QUALITY MIX BLENDER
Whole medical range is available on
Quality Mix provides a continuous and precise mixture of medical Air and Oxygen for infants, children and adults. The exact O2 concentration is easy to adjust and an optional bleed switch enables very low gas consumption, even with flow rates of less than 3 l/pm.
- The oxygen concentration can be set easily and accurately between a range of 21 % and 100 %.
- A Pressure Alarm sounds if there is any interruption or pressure drop in the gas supply system.
- A robust metal housing protects the device from damage, especially under tough clinical conditions.
- Alarm limits can be set to ensure the corresponding oxygen concentration.
- The delivered oxygen concentration can be displayed on the optional oxygen monitor.
735100001233 Quality Mix Blender leaflet EN
TECHNICAL DATA
Blender Type: | High Flow | Low Flow |
Main flow outlet (l/min): | 15-120 | 3-30 |
Auxillary flow (l/min): | 2-100 | 0-30 |
Max. combined flow - both outlets (l/min): | >120 | >30 |
Bleed flow at 3,4 bar (l/min): | <13 | <3 |
Emergency flow (in the event that Air or O² supply failing) (l/min): | <85 | <15 |
Pressure drop at inlet pressure from 3,1 to 5,2 bar with a flow rate of 10l/min / 30l/min at 60%FI O2 | <0,21 | <0,14 |
Alarm sounds when supply pressure drops (bar): | ||
Without regulator module: | on 3,3 | off 4,2 | on 3,3 | off 4,2 |
With regulator module: | on 2,3 | off 3,2 | on 2,3 | off 3,2 |
Alarm volume: | >80dB at a distance of 30 cm | |
Setting range for oxygen concentration: | 21-100 % | |
Gas inlet pressure: | 4,5±0,5 bar: air and O2 differential should be with 0,7 bar | |
Accuracy of mixed gas (FIO2): | ± 3% oxygen | |
Connection types: | DISS inlets and outlets for O2 and/or NIST inlets for air and O2 | |
Dimensions (LxWxH): | 130×165×122 mm | |
Weight: | 1600g | |
Operating temperature: | +5°C to + 40°C | |
Storage temperature: | -20°C to + 60°C | |
Humidity: | Max 95% non-condensing humidity |
* All blenders will be delivered in a carton box 42×42×32 cm (L×B×H) with compartments for the manuals and all ordered accessories.