GCE GROUP ANNOUNCES REBRANDING FIND OUT MORE close
Home » MEDIUNIT AFNOR TYPE

MEDIUNIT AFNOR TYPE

Group Number: 3655

Whole medical range is available on

Medical terminal units provide a quick and easy connection of hospital ward gas equipment to the hospital gas source. The type of medical gas outlets are decided by national standards in each country and sometimes from local requests in each hospital. GCE complies with ISO 7396 and national installation standards with secure products where every product is fully tested in production. Our Medical gas outlets are in accordance with the ISO EN 9170-1 international standard.

  • All functional components are from brass
  • Simple installation
  • Fast connection and disconnection
  • Air–800 with a parking position
  • Designed for the medical environment, small size and easy to clean
  • Complies with colour coding and description by standard
  • Maintenance valve with filter
  • Recessed and exposed versions
  • Bed head installation versions (customized solution on request)

MediUnit Afnor type EN

TECHNICAL DATA

Gases: O2, N2O, Air, Air-800, CO2, N2, VAC
Dimensions:  
Height: 73 mm
Width: 73 mm
Depth: 63 mm
Working pressure: 4–5 bar (breathing gases)
  7–10 bar (instrumental gases)
  (-0,4)–(-0,9) bar (vacuum)
Maximum pressure: 20 bar
Regulatory status:

Complies with Medical Directive 93/42/EEC

Complies with EN ISO 7396-1 
(Central Gas Supply systems)

Complies with EN ISO 9170-1 
(Terminal units)

 

Complies with NF S 90-116
(Afnor gas specific connections)

 

Complies with FD S 90-119
(Afnor Air-800 gas specific connection) 

 

present HTM 02-01

Technical Datasheets Top

Download Technical Datasheets