GCE集团宣布重塑品牌 点击了解更多 close
主页 » LINE REGULATOR

LINE REGULATOR

组号: 3611

Whole medical range is available on

A stabilizer is a pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
The stabilizer makes it possible to distribute gas with different pressure to department and buildings in the hospital area. In some cases it is needed to deliver a higher pressure from the main gas manifold to compensate for small pipe dimensions. In those cases the Stabilizer shall be mounted as close as possible before the first terminal unit, to ensure a correct pressure to the patient.

FOR A COMPLETE LINE REGULATOR ADD:

  • Plastic cover for locking
  • Alarm unit (included if ordered together with HP unit)
Item No.  Denomination Gas type
Inlet*
0727333 LINE REG O2, N2O, Air, CO2, N2 LH
K141621 LINE REG O2 – AFNOR LH
K141631 LINE REG O2 – AFNOR RH
K141622 LINE REG N2O – AFNOR LH
K141632 LINE REG N2O – AFNOR RH
K141623 LINE REG Air – AFNOR LH
K141633 LINE REG Air – AFNOR RH
K141629 LINE REG Air-800 – AFNOR  LH
K141639 LINE REG Air-800 – AFNOR  RH
K141624 LINE REG N2 – AFNOR  LH
K141625 LINE REG CO2 – AFNOR  LH

* LH = inlet from left side; RH = inlet from right side

735100001553 Line Regulator leaflet EN

TECHNICAL DATA

Gases: O2, Air, Air-800, N2, N2O, CO2, (all medical gases)
Nominal flow: 40 m3/h
Inlet nominal pressure: 16 bar (1600 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–10 bar)
Inlet connection: G1/2"M+ soldering piece pipe ∅ 12 mm
Outlet connection: G1/2"M+ soldering piece pipe ∅ 12 mm
Pressure sensors: Optional (Pressure switches; Transmitters
0–50 mV; 4–20 mA
Emergency QC inlet: Optional QC by national standards

Regulatory status:

 

 

 

Complies with Medical Devices Directive 93/42/EEC

Complies with EN ISO 7396-1
(Central Gas Supply Systems)

Complies with EN 60601-1-2
(Electromagnetic compatibility)

present HTM 02-01

技术参数 顶部

下载技术参数