GCE集团宣布重塑品牌 点击了解更多 close
主页 » GAS MANIFOLD MM90 - HP UNIT FULLY AUTOMATIC

GAS MANIFOLD MM90 - HP UNIT FULLY AUTOMATIC

组号: 3611

Whole medical range is available on

The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm). As 2nd stage is recommended to use a stabilizer. The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted. At that point the supply will switch to the reserve bank and the empty bank can be replenished. The object gives uninterrupted gas supply. Gas Alarm is a standard accessory. The gas alarm gives visual and audible indication.
It surveils and the alarm sounds when the following happens:

1.  Change operation side / Leaking on reserve side
2.  High operation pressure
3.  Low operation pressure
4.  Empty possision (High/Low distribution pressure when connected to a Stabilizer)

The Gas alarm is able to communicate with other equipment through relays.

MM90 INCLUDES THE FOLLOWING COMPONENTS:

  • MM90 HP unit Manifold
  • Gas alarm
  • Purge valves
  • HP filters

FOR A COMPLETE MM90 HP UNIT MANIFOLD ADD:

  • Collecting pipe set (high pressure valves and non-return valves/high pressure components)
  • Cylinder retaining brackets (included in gas cylinder collecting pipe set)
  • High pressure hoses with safety wire
  • Plug for close collecting pipeline
  • Gas name sign
  • Stabilizer

735100000439 MM90 HP unit fully automatic leaflet EN

Item No.  Denomination Gas PRV Alarm
0732322 MM90 – HP unit AUTO 2×1 O2 Standard Alarm TMA6703
0732323 MM90 – HP unit AUTO 2×1 Air Standard Alarm TMA6703
0732324 MM90 – HP unit AUTO 2×1 N2O, CO2 Standard Alarm TMA6703
0727308* MM90 – HP unit AUTO 2×1 O2, N2O, Air, CO2, N2 Standard
0727309 MM90 – HP unit AUTO 2×1 O2 Standard Pressure switch
0727310 MM90 – HP unit AUTO 2×1 Air Standard Pressure switch
0727311 MM90 – HP unit AUTO 2×1 N2O, CO2 Standard Pressure switch

* basic version without electronic sensors

TECHNICAL DATA

Gases: O2, Air, N2, N2O, CO2,  (all medical gases)
Nominal flow: 90 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 9 bar (setting range 9-15 bar)
Inlet connection: W21,8×1/14"M
Outlet connection:

G3/4"F + soldering piece pipe ∅ 22 mm

Intermediate pressure relieve valve: 16 bar
Pressure relieve valve pipe dimension: ∅ 10 mm
Purge valves connection: W21,8×1/14"M + soldering piece pipe ø 10 mm

Regulatory status:

 

 

 

Complies with Medical Devices Directive 93/42/EEC

Complies with EN ISO 7396-1
(Central Gas Supply Systems)

Complies with EN 60601-1-2
(Electromagnetic compatibility)

present SIS HB 370 and HTM 02-01

技术图纸 顶部

下载技术图纸