GAS MANIFOLD MC150 - STABILIZER
组号: 3611Whole medical range is available on
MC150 - STABILIZER is a second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
MC150 - STABILIZER is a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MC80, MM90 or liquid oxygen tank (LOX)). When there is a signal for pressure deviation in relation to the alarm settings, the alarm can easily be displayed on a Gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with either pressure transmitter 4-20 mA or with contact gauge.
The gas reduction unit MC150 - STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
FOR A COMPLETE MC150 STABILIZER MANIFOLD ADD:
- Alarm unit (included if ordered together with HP unit)
Item No. |
Denomination | Gas | PRV | Alarm |
325397706 | MC150 STAB | O2 | With pull-ring | Contact gauge |
325397707 | MC150 STAB | O2, Air, N2 | With pull-ring | 4–20 mA |
735100000439 MC150 Stabilizer leaflet EN
TECHNICAL DATA
Gases: | O2, Air, N2, N2O, CO2, (all medical gases) |
Nominal flow: | 360 m3/h |
Inlet maximal pressure: | 20 bar (2000 kPa) |
Outlet nominal pressure: | 4,5 bar (setting range 0,5–6 bar) |
Inlet connection: | 2× G1 1/2"F + soldering piece pipe ∅ 35 mm |
Outlet connection: | 2× G1 1/2"F + soldering piece pipe ∅ 35 mm |
Pressure relieve valve: | 6,8 bar |
Pressure relieve valve pipe dimension: |
∅ 35 mm |
Regulatory status: |
Complies with Medical Devices Directive 93/42/EEC Complies with EN 60601-1-2 present SIS HB 370 and HTM 02-01 |